Sensor Specifications | ||
Sensor | Polarographic, Gold/Silver, PTFE membrane | |
Range | Chlorine: 0.05 to 20 ppm (Standard Range) 0.01 to 5 ppm (Low Range) pH: 0 to 14 pH |
|
Temperature | 32° to 122°F (0° to 50°C) | |
Flow | 38 L/hr to 300 L/hr (10 gal/hr to 80 gal/hr) | |
Wetted Materials | PVC, PP, PVDF, PTFE, Glass, 316 SS | |
Process Connections | Input 1/4″ barb fitting, drain 3/4″ barb fitting | |
Response Time | T90 in approximately 2 minutes | |
Electrolyte Life | Up to 12 months |
Analyzer Specifications | ||
Measurements | Chlorine: 0.05 to 20 ppm (Standard Range) 0.01 to 5 ppm (Low Range) pH: 0 to 14 pH |
|
pH Compensation | 5-10 pH | |
Display | 2.75″W x 1.5″H (128 x 64 pixels) LCD, Black on Grey background, Blue on White background with LED back-light |
|
Enclosure | IP65, weatherproof, 1/2 DIN 5.7″L x 5.7″W x 3.5″D |
|
Output | (1) 4-20 mA for Chlorine, set to sensor range (1) 4-20 mA for pH, set to 0-14 pH |
|
Input Power | Standard: 100-240 VAC, 50/60 Hz, 4W, protected with 250V, 1A, Slow Blow Fuse Optional: 24 VDC (18-36 VDC @ 250 mW minimum) |
|
Alarm Relay Ratings | Three (3) SPDT, Form C, 250V, 10 Amp | |
Options | HART Protocol Spray Cleaning (100V or 240V) High Flow CFD NEMA 4X System enclosure with clear or opaque door |
|
Accessories | Photometric Tester Kit, Calibration/Validation Testing Strips Membrane Replacement Kit Sensor Polish Kit |