VD6S2230 with VSP63MA4 – Short Path Distillation Systems
● 1000 to 1e-4 mbar
he Thyracont VD6 universal vacuum indicator and controller is designed for panel mounting. A relay monitor on the front panel allows continuous operation control of the two switch points.
Complete Set consisting of:
- VD6S2230 vacuum display and control unit
- VSP63MA4 vacuum transducer
- W0509002 measuring cable
Short Path Distillation Systems
Thermally sensitive product mixtures can be separated by means of the so called short path distillation. As the working pressure is decreased into the fine vacuum range this distillation is carried out at very low temperature.
Challenge
By means of a wiper system the raw material to be processed is dispersed on a tube wall which is heated from the outside. Low-boiling components evaporate and are recollected at an internal condenser. The process pressure, which typically lies in the range 1e-3 mbar to 1 mbar, must be kept constant and recorded for quality assurance.
Solution
The vacuum display and control unit VD64P is used for monitoring and controlling process pressure. With isolated relay switch points a vacuum valve and, if necessary, a venting valve can be controlled. In combination with the included VSP63 Pirani transducer the required vacuum control can be realized in a reliable and cost-effective way. The serial USB interface of the VD6 is used for recording the pressure data during the whole process.
Benefits at a Glance
- Complete measurement set with Analogline Pirani transducer (VSP63MA4), measuring cable and controller
- VD6 controller with 1 measuring channel 2 potential-free relay switches
- EMI compatible
- Intuitive operation
| Weight | 4.0 kg |
|---|---|
| Dimensions | 28 x 28 x 25 cm |
| Suitable vacuum gauge | VCC200MA4, VSC43MA4, VSP63MA4 |
| scanning rate | 2 Hz (0,5 s) |
| Scanning Rate | 20 Hz (50 ms) |
| Voltage supply | 230 VAC, 50/60 Hz |
| Power consumption | Max. 3.5 W |
| Display | LED, 13 mm, max. 4 1/2 digits, display unit mbar (Torr on request) |
| Operating temperature | +5…+50 °C |
| Storage temperature | -20…+60 °C |
| Messeingänge | 4 – 20 mA, SubD, 9pole, female |
| Switching output | 2 x relay contact, potential free change-over switches, Phoenix strip terminal 3-pole each: 2 A / 250 VAC, 1 A /30 VDC (normal closed); 5 A / 250 VAC, 5 A / 30 VDC (normal open) |
| recorder output | 0-1 V or 0-10 V |
| Serial interface | Mini-USB, Typ B, Virtual Com Port Protokoll |
| Protection class | IP20 |
| Weight | 410 g |
| Brands |
Thyracont Vacuum Instruments |
|---|
Related products
VD84 – Vacuum Insulation of Liquid Gas Tanks
Thyracont Vacuum Instruments